Probe Specifications: |
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
|
(general-purpose coating) |
(high-sensitivity for low levels of O2) |
(for hydrocarbon environments) |
O2% range (at 1 ATM): |
0-100% |
0-10% |
0-20.9% |
DO range (ppm at 1 ATM): |
0-40 ppm |
0-4 ppm |
0-8 ppm |
Temperature range: |
-50-+80 oC for probes |
-50-+80 oC for probes |
-50-+60 oC |
0-60 oC for patches |
0-60 oC for patches |
NA |
O2% resolution: |
0.01-0.05% |
.001% |
0.01-0.05% |
DO resolution (at room temp): |
4-20 ppb |
0.4-2 ppb |
4-20 ppb |
O2% accuracy: |
5% of reading |
5% of reading |
5% of reading |
DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
Min. detectable level in gas: |
0.01% - 0.05% |
0.001% - 0.01% |
0.01% - 0.05% |
Response time: |
< 1 second in gas |
< 1 second in gas |
< 1 second in gas |
~30 seconds with overcoating in gas |
~30 seconds with overcoating in gas |
~45 seconds in pure water |
|
~45 seconds in pure water |
~45 seconds in pure water |
|
|
Patch Specifications: |
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
02% range (at 1 ATM): |
0-100% |
0-20.9% |
0-100% |
DO range (ppm at 1 ATM): |
0-40 ppm |
0-8 ppm |
0-40 ppm |
Temperature range: |
-20 to +60 oC for patches |
-20 to +60 oC for patches |
-20 to +60 oC for patches |
02% resolution (20 sex averaging): |
0.05% |
0.01% |
0.05% |
DO resolution (at room temp): |
20 ppb |
4 ppb |
20 ppb |
02% accuracy: |
5% of reading |
5% of reading |
5% of reading |
DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
Min. detectable level: |
0.1% O2 |
0.01% O2 |
0.1% O2 |
Min detectable level in water (at room temp): |
40 ppb |
4 ppb |
40 ppb |
Response time: |
<1 second in gas |
<1 second in gas |
<1 second in gas |
|
~30 seconds with overcoating in gas |
~30 seconds with overcoating in gas |
|
|
~45 seconds in pure water |
~45 seconds in pure water |
~45 seconds in pure water |
|
Sterilization Guide: |
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
Form factor: |
Probes/Patches |
Probes/Patches |
Probes |
Autoclavable: |
No |
No |
Yes |
Calibration: |
|
|
Single point reset required after sterilization |